Base thermal cycler
|
DNA Engine
|
Sample capacity
|
96-well PCR plate (low-profile) / 12 x 0.2 ml 8-tube strips (low-profile)
|
Sample volume
|
10–100 µl (20 µl recommended)
|
Dimensions (W x D x H)
|
34 x 47 x 60 cm
|
Weight
|
29 kg
|
Fluorescence excitation range
|
470–505 nm
|
Fluorescence detection range
|
Channel 1: 523–543 nm
Channel 2: 540–700 nm
|
Linear dynamic range of starting copy number
|
Up to 10 orders of magnitude
|
Detection limit of template starting copy number
|
Down to single initial template copy detection
|
Gradient Specifications
|
Gradient range
|
30–105°C
|
Temperature differential range
|
1–24°C
|
Gradient accuracy
|
±0.3°C of target at end columns within 30 sec of arrival at programmed target, NIST–traceable
|
Temperature uniformity
|
±0.4°C, in column, well-to-well, within 10 sec of target attainment
|
Gradient calculator accuracy
|
±0.4°C of actual column temperature (NIST–traceable)
|
Computer System Requirements (Computer NOT Included With System)
|
Model
|
Desktop tower computer
|
PCI card slot dimensions
(L x H)
|
Full-size card, 280 mm (11 in) x 106 mm (4.2 in) / short card, 174 mm (6.9 in) x 106 mm (4.2 in)
|
Operating system
|
Windows XP Professional
|
Storage capacity
|
40 GB hard drive
|
Memory
|
512 MB RAM
|
Port
|
Communication (COM)
|
Supplied With System
|
Data acquisition board
|
National Instruments PCI-6036E, 200 x 10³ samples/sec
|
Note: Specifications subject to change without notice.
|